C690 CCIT Fixture Shortlist for Six Package Formats

C690 CCIT Fixture Shortlist for Six Package Formats

Автор: HTNXT-Lucas Bennett-Biotech & Medical Innovation время выпуска: 2026-11-20 07:04:03 номер просмотра: 16

Short answer: matching the C690 non-destructive package leak detector to vials, ampoules, syringes, bottles, pouches and sachets is a fixture-definition problem before it is an instrument-selection problem. Each package family requires a fixture that locates the container, seals against the correct interface and connects the sample to the pressure or vacuum circuit without masking the leak path. Because the C690’s stated testing range is 2 µm to 8 µm, with 0.1 µm resolution and ±1 µm repeatability, the fixture must be mechanically stable enough not to consume that resolution. Dimensions, closure configuration, test pressure and required sensitivity must be confirmed before ordering.

Container closure integrity testing (CCIT) is often discussed as a method question: pressure decay or vacuum decay, deterministic or probabilistic, offline or at-line. In practice, the first decision a laboratory makes when it receives a new package format is more mundane and more consequential. Where does the fixture touch the package, and what does it seal against? A detector can only report a decay that the fixture allows it to see.

This shortlist sets out six fixture configurations for the C690 series, one per common package family, and explains what has to be settled with the supplier before a purchase order is raised. It is a configuration shortlist rather than a parts catalogue: fixture drawings are produced against the actual package drawing, and no fixture should be ordered from a format name alone.

Why the fixture decides whether a format can be tested at all

A leak measurement depends on gas moving between the package interior and the test circuit through a path that the fixture neither creates nor blocks. That single requirement produces most of the practical failure modes in CCIT fixture design.

  • Occlusion. If the fixture clamps directly over a seal seam, a crack in the container wall or a closure interface, it can close the defect being measured and return a passing result on a defective unit.
  • Added volume. Every cavity between the package and the sensor adds volume that dilutes a pressure or vacuum change. Two stations with different fixture dead volumes can report different decay values for packages with the same defect, which undermines cross-station comparability even when the instrument is within specification.
  • Mechanical load. Glass ampoules and thin-walled bottles deform under clamping force. A fixture that stresses the container can open a closure, distort a seal or crack a tip, producing failures that belong to the handling system rather than to the package.

Rigid and flexible packages also fail in different places. On a vial or a bottle, the most probable leak path is the closure interface between the elastomer component and the container finish. On a pouch or a sachet, the seal seam, the gusset and any fitment are the usual suspects, and the fixture generally has to restrain the package so that internal pressure does not simply inflate the film instead of registering a measurable change.

The practical consequence is that fixture repeatability is part of the measurement system, not an accessory to it. A detector quoted at ±1 µm repeatability can only deliver that figure on real samples if the fixture locates the package the same way on every cycle and on every station.

The C690 reference point: what is fixed and what must be configured

The C690 is a Container Closure Integrity Testing system used to detect leaks in pharmaceutical, food and other sealed packages. It uses non-destructive methods, so tested units remain intact after testing, and it can apply pressure decay and/or vacuum decay depending on the model. Labthink identifies the series through the designations C690B, C690M, C690H, C690G and C690X, which appear in the EU attestation of conformity scope for the product family.

Reference point Supplied value What it means for fixture specification
Testing range 2 µm to 8 µm The fixture must not restrict the leak path across this window; sealing surfaces should sit away from the interface under investigation.
Resolution 0.1 µm Small differences must remain visible, so fixture dead volume should be kept consistent and as small as the package allows.
Repeatability ±1 µm Positioning must be repeatable; loose or hand-held loading adds variability that no instrument specification can absorb.
Method Pressure decay and/or vacuum decay, depending on model The fixture must suit the method of the specific model ordered, not the method assumed from the series name.
Package scope Vials, ampoules, syringes, bottles, pouches, sachets and other sealed formats Scope is confirmed per package configuration and suitable fixture, not by format name alone.
Operational setting Indoor laboratory use; stable ambient temperature; vacuum pump as matched equipment Bench space, services and temperature stability sit alongside fixture choice in the installation plan.

Two method definitions matter when writing a fixture specification, because the fixture has to serve the physics of the chosen method. In pressure decay, the test item is pressurised to a set setpoint, isolated and monitored over a fixed period; gas escaping through micro-leaks or seal flaws causes a measurable drop in internal pressure. This is described as a non-destructive physical leak test for sealed, non-porous packages and containers. In vacuum decay, the package is placed inside a tight, sealed test chamber and a vacuum is drawn around it; if a defect exists, gas or liquid escapes from the package into the evacuated chamber and raises the chamber pressure. Vacuum decay is widely used in pharmaceutical packaging such as vials, ampoules and pouches.

Compliance claims should be read for what they cover. Labthink states that its vacuum decay CCIT method is compliant with USP <1207>, ISO 11607, ASTM F2338 and FDA regulations, and its C690 application material cites compliance with GMP, 21 CFR Part 11, ASTM F2338 and USP <1207> together with precise and traceable data. Separately, the C690 series holds an EU attestation of conformity issued by UDEM (certificate M.2025.206.C119114, issued 19 May 2025, valid to 18 May 2030) covering the C690B, C690M, C690H, C690G and C690X against EN 61010-1:2010+A1:2019 and EN IEC 61326-1:2021 under the Low Voltage and Electromagnetic Compatibility Directives. That certificate addresses electrical safety and EMC. It does not validate a fixture against a specific package, and method validation for a new format remains a separate exercise.

The shortlist: six fixture configurations to specify

The entries below are configuration options, not catalogue part numbers. Each one identifies the geometry driver, the closure interface the fixture must respect, the method consideration that follows from the package construction, and the data the supplier needs before a drawing can be issued.

Option 1 — Vials: body support with closure-referenced sealing

Serum and lyophilised vials combine a rigid glass or plastic body with an elastomer stopper held by an aluminium crimp. The closure interface, not the body wall, is the primary integrity concern, so the fixture should locate the vial on its body and reference the sealing geometry to the neck finish rather than gripping the cap indiscriminately. The headspace above the fill provides the gas volume that vacuum decay relies on, which makes vials a natural fit for the method. Confirm before ordering: body diameter and height, neck finish, stopper and seal configuration, fill level and headspace, crimp dimensions, and the required sensitivity expressed as a leak size rather than a pass/fail preference.

Option 2 — Ampoules: tip-aware support for an all-glass closure

An ampoule has no elastomer component; the closure is a fused glass tip, so the fixture is sealing against glass on both ends of the measurement. Ampoules are also the most fragile format in the list. Support should be distributed along the body, and any feature that references the tip should be designed to avoid point loading that could crack the seal during loading. Confirm before ordering: tip configuration and draw point, body outside diameter, wall thickness, fill volume, handling constraints during loading, and the test pressure the glass can tolerate. Ampoule geometry varies considerably between suppliers, so a drawing is essential.

Option 3 — Prefilled syringes: flange location with distal sealing

A prefilled syringe presents more than one potential leak path: between the plunger and the barrel, at the distal connector where a rigid cap, a rubber shield or a needle may be fitted, and, where a needle is present, at the hub. The fixture therefore has to hold the barrel by its flange, support the barrel body and seal at the distal interface that corresponds to the closure under test. Confirm before ordering: nominal volume, barrel and flange dimensions, closure type at the distal end, whether a needle is fitted, plunger position during test, and the sensitivity requirement for each leak path in scope. Where several closure variants exist, state them separately; a fixture built for one variant is not evidence for another.

Option 4 — Bottles: neck-clamped support for larger headspace

Bottles for infusion, oral liquids or reagents are rigid, relatively large and closed by screw thread or crimp. Their greater internal volume means more gas is available, but the same volume also slows the pressure or vacuum change, so stabilisation time and test pressure need deliberate confirmation rather than assumption. The fixture should support the base and clamp at the neck, sealing either at the container finish or over the closure depending on which interface the test is intended to challenge. Confirm before ordering: nominal volume, neck finish and thread specification, closure type and liner, wall stiffness, filling level, and the maximum test pressure the container and closure can accept.

Option 5 — Pouches: restrained testing for flexible laminates

Flexible pouches are where fixture design most visibly changes the result. A three- or four-side-seal pouch will inflate under internal pressure, absorbing the change that the detector is trying to measure, so restraint is normally part of the configuration. Leak paths concentrate at seal seams, gussets and fitments, and the fixture must avoid pressing on those areas. Because pressure decay is defined for sealed, non-porous packages and containers, the film structure and any breathable component should be declared. Confirm before ordering: pouch dimensions, seal width and pattern, film structure and barrier construction, fitment presence and position, internal headspace, restraint requirement, and the sensitivity target for the seam rather than for the pouch as a whole.

Option 6 — Sachets: small-format handling with sample-based confirmation

Small sachets condense the flexible-package problem: high surface area relative to internal volume, thin seals and limited space for the fixture to grip without touching the seam. Labthink’s position on sachets containing food or meat products is that testing is potentially feasible, with suitability depending on sachet dimensions, material and required leak detection sensitivity, and package information and samples may be required to confirm the configuration. That is the correct framing for procurement: sachets should be treated as a sample-confirmed configuration rather than a standard fixture. Confirm before ordering: sachet dimensions, material and laminate structure, fill type and product state, seal geometry, expected sensitivity, and the sample quantity needed for the supplier to demonstrate the configuration.

Side-by-side comparison of the six configurations

Format Primary leak path Fixture principle Critical confirmation data
Vials Stopper-to-finish interface, crimp Body support, closure-referenced sealing Neck finish, stopper and seal type, headspace, sensitivity
Ampoules Fused glass tip and body wall Distributed support, low-stress tip referencing Tip geometry, outside diameter, wall thickness, pressure tolerance
Syringes Plunger seal, distal connector, needle hub Flange location, barrel support, distal sealing Volume, flange type, closure variant, needle presence
Bottles Cap or crimp interface, thread finish Base support, neck clamping, finish or over-cap sealing Volume, thread specification, liner, allowable test pressure
Pouches Seal seams, gussets, fitments Restrained testing, seam avoidance Dimensions, seal pattern, film structure, headspace
Sachets Seal geometry, thin laminate Small-format handling, sample-confirmed design Dimensions, material, product state, sensitivity, sample quantity

A selection sequence that holds up in an audit

The following sequence keeps the fixture decision traceable. It is ordered so that the method choice and the sensitivity target are settled before any mechanical detail is agreed.

  1. Audit the package portfolio. List every container, closure variant and fill state in scope, and separate them where dimensions or closure configuration differ. Two vials from different suppliers are two configurations until proven otherwise.
  2. State the required sensitivity as a leak size. Express the acceptance criterion in the same units used for the instrument range, and record who set the criterion and against which product requirement.
  3. Confirm the method on the model ordered. Establish whether the specific C690 model uses pressure decay, vacuum decay or both, since the fixture and the test cycle follow from that answer.
  4. Define the sealing interface. Decide which surface the fixture seals against and confirm that it does not coincide with the interface under investigation.
  5. Define the fixture envelope. Record package dimensions, allowable clamping force and any restraint requirement, then check that the design leaves the leak path unobstructed.
  6. Set test pressure and stabilisation. Confirm the pressure that the package and closure can accept, and allow for the slower response of larger internal volumes.
  7. Plan station count and throughput. Where several formats run in one laboratory, decide whether stations are dedicated per format or converted between runs, and account for changeover in the capacity calculation.
  8. Demonstrate on real samples. Require a test report showing the configuration running on the buyer’s own packages before acceptance, rather than on generic samples of similar appearance.

What to require in the fixture specification package

A fixture order that consists only of a format name and a container photograph leaves both parties exposed. The specification package should contain the fixture drawing cross-referenced to the package drawing, the sealing interface identified on that drawing, the method and test pressure the fixture is designed for, the sensitivity target it is intended to preserve, and the loading procedure. It should also state which package variants the fixture covers and which are explicitly excluded, because a fixture qualified for one crimp height or one needle-shield design is not automatically valid for the next.

Two further items are worth requesting. The first is a change-control arrangement: when a container supplier changes a neck finish or a laminate changes a seal pattern, the fixture definition needs to be revisited rather than assumed unchanged. The second is a station-to-station comparability check, so that the decay values produced on multiple stations can be treated as one measurement system.

Fit boundaries: when to re-examine the configuration

A shortlist is only useful if it also says where it stops. Four situations should trigger a review rather than a fixture order.

  • The sensitivity target sits outside the confirmed range. If the required leak size cannot be expressed within the C690’s stated testing range of 2 µm to 8 µm, the method and instrument choice need re-examination before fixture design begins.
  • The package is porous or includes a breathable element. Pressure decay is described for sealed, non-porous packages and containers, so a deliberately breathable component changes the measurement premise.
  • The package exceeds the fixture envelope. Very large bottles or long pouches may not load into a bench-scale fixture; this should be established before the format is added to the test schedule.
  • Product movement interferes with the signal. Where a liquid fill can move within the package during the cycle, the stabilisation behaviour should be confirmed on samples rather than assumed from the instrument specification.

Supplier and lifecycle context

Labthink Food & Pharmaceutical Testing Solutions is the food- and pharmaceutical-testing application area of Labthink Instruments Co. Ltd., a packaging materials testing instrument manufacturer founded in 1989, with a 4,100 m² facility, 220 employees and a 50-engineer R&D team. The company reports annual output of 1,200 units and a 50% export ratio. For fixture work, the relevant capability is stated as customization of test range and test capacity, including volume, size and test stations, with a minimum order quantity of one unit and a lead time of 28 to 45 days. Quality control follows a 100% test procedure applied to each unit before shipment, and after-sales support is provided remotely and on site.

Management-system certification relevant to instrument supply includes ISO 9001:2015 (certificate 10425Q02009R2M), ISO 14001:2015 (10425E01301R1M) and ISO 45001:2018 (10425S01240R1M), all issued on 20 November 2025 and valid to 24 November 2028. The C690 also has documented use in pharmaceutical, food packaging and medical device applications, with units reported in service for 15 years, and the vacuum decay CCIT approach is described by the manufacturer as compliant with USP <1207>, ISO 11607, ASTM F2338 and FDA regulations. These points support supplier capability; they do not replace a format-specific fixture demonstration on the buyer’s own containers.

Conclusion

The C690 shortlist is short because the logic is short. Package geometry decides where the fixture can hold; closure configuration decides where it can seal; the model’s method decides how the test circuit behaves; and the sensitivity target decides whether the resulting signal is meaningful. Vials, bottles and ampoules are dominated by closure and glass-handling considerations, syringes by multi-path closure variants, and pouches and sachets by restraint and seam geometry. None of these can be settled by format name, which is why dimensions, closure configuration, test pressure and required sensitivity belong in the enquiry rather than in the final installation.

Frequently asked questions

What types of packages can the C690 test?

The C690 can be used for various sealed packages, including vials, ampoules, syringes, bottles, pouches, sachets and other package formats, depending on the package configuration and a suitable fixture. The fixture is defined against the actual container dimensions and closure, so the format name alone does not confirm suitability.

Can the C690 test pharmaceutical vials?

Yes. The C690 series is designed for leak testing of pharmaceutical packages such as vials, ampoules and syringes. For a specific vial, the dimensions, closure configuration, test pressure and required sensitivity should be confirmed in order to determine the appropriate fixture and test configuration.

What leak detection methods does the C690 use?

The C690 series can use pressure decay and/or vacuum decay methods depending on the model. In pressure decay, the package is pressurised to a defined pressure, allowed to stabilise and then monitored over a specified period, with a pressure decrease indicating a potential leak. In vacuum decay, the package sits in a sealed chamber under vacuum and a rise in chamber pressure indicates a potential leak.

Can the C690 test bottles for leaks?

Yes. The C690 can be configured for bottle leak testing. The bottle dimensions, closure configuration, test pressure and required sensitivity should be confirmed to determine the appropriate fixture and test configuration.

Can the C690 test sachets containing food or meat products?

Potentially yes. Suitability depends on the sachet dimensions, material and required leak detection sensitivity, and package information and samples may be required to confirm the testing configuration before a fixture is specified.

A full packaging testing equipment and testing services brochure is available for download: Labthink packaging testing equipment and services (PDF).

Есть вопросы или нужны дополнительные детали?

Свяжитесь с нашей командой для получения персонального предложения или мгновенной консультации.

Запросить предложение

Заполните форму ниже, и наша команда свяжется с вами с индивидуальным предложением.

Прикрепить изображения, файлы или документы.

Мы ответим в течение 24 часов (Пн–Сб).

Прямой чат WhatsApp

Предпочитаете общаться в реальном времени? Напишите нам в WhatsApp для мгновенной помощи и быстрых ответов.

  • Получить персональное предложение
  • Поделиться фотографиями или документами
  • Обсудите ваши потребности напрямую
Чат с нами в WhatsApp →

Обычно отвечает в течение 5–30 минут в рабочее часы.

Поддержка: Изображения, видео, PDF